AI – Axis Interface (layer 5)
0x15000001 | System Error calling a RTOS function |
0x1500?010 | Output data of previous clock not written |
0x1500?011 | Excessive change in position |
0x1500?012 | Excessive lag |
0x1500?013 | Overspeed |
0x1500?014 | PHM communication error |
0x1500?015 | PHM datum error during movement |
0x1500?016 | PHM not connected |
0x1500?017 | PHM control code error |
0x1500?018 | PHM transmission error |
0x1500?019 | PHM invalid data |
0x1500?01A | PHM datum error (not seated) |
0x1500?01B | PHM obstruction |
0x1500?01C | PHM overload |
0x1500?01D | General hardware error |
0x1500?01E | Transducer error, unknown code meaning |
0x1500?01F | Drive error, unknown code meaning |
AI – Axis Interface (layer 5) 0x1500002x
0x1500?020 | Weight Balance routine timed out after a PH rotation |
0x1500?021 | Drive error: IxT |
0x1500?022 | Drive error: voltage saturation |
0x1500?023 | Drive error: tacho speed too high |
0x1500?024 | Drive error: command refresh timeout |
0x1500?025 | Drive error: VP overvoltage |
0x1500?026 | Drive error: VP undervoltage |
0x1500?027 | Drive error: servodrive overtemperature |
0x1500?028 | Drive error: VP overcurrent |
0x1500?029 | Drive error: brake resistor failure |
0x1500?02A | Drive error: MOSFET auxiliary voltage failure |
0x1500?02B | Drive error: AD converter failure |
0x1500?02C | Drive error: invalid DSP firmware |
0x1500?02D | Drive error: invalid DSP params |
0x1500?02E | Drive error: invalid SET_PARAM command received |
0x1500?02F | Drive error: RAM check failed |
AI – Axis Interface (layer 5) 0x1500003x
0x1500?030 | Drive error: CPLD check failed |
0x1500?031 | Drive error: short circuit on a digital output |
0x1500?032 | Drive error: failure programming the transducer offset E2POT |
0x1500?033 | Drive error: Hardware error no Ready Pwm |
0x1500?034 | Drive error: signals GFAIL and READYPWM not coherent |
0x1500?035 | Drive error: signals NINH and POWON not coherent |
0x1500?036 | Drive error: signals NINH and BRKN not coherent: failure detected on KP relay |
0x1500?037 | Drive error: unknown event received |
0x1500?038 | Transducer error: TTL alarm signal detected (TRASER) |
0x1500?039 | Transducer error: TTL missing signal (FLET) |
0x1500?03A | Transducer error: edge separation |
0x1500?03B | Transducer error: speed mismatch |
0x1500?03C | Transducer error: amplitude error |
0x1500?03D | Transducer error: phase error |
0x1500?03E | HF instability |
0x1500?03F | Excessive dual disalignment |
AI – Axis Interface (layer 5) 0x1500004x
0x1500?040 | Force limitation threshold exceeded |
0x1500?041 | Force limitation algorithm intervention |
0x1500?042 | THD answer timeout error |
0x1500?043 | No HP-THD probe detected |
0x1500?044 | Invalid THD probe parameters |
0x1500?045 | Error during THD probe configuration |
0x1500?046 | Failure opening a file |
0x1500?047 | Error during THD controller configuration |
0x1500?048 | PHM firmware update required to use thdConfig parameters |